MWexperimental | 6 | kDa |
MWexpected | 6 | kDa |
VPorod | 11 | nm3 |
log I(s)
3.30×10-3
3.30×10-4
3.30×10-5
3.30×10-6
|
s, nm-1
|
|
|
|
|
Synchrotron SAXS
data from solutions of
G-quadruplex DNA from Hepatitis B virus (mutant)
in
20 mM HEPES, 100 mM KCl, 1 mM EDTA, pH 7.5
were collected
on the
B21 beam line
at the Diamond Light Source storage ring
(Didcot, UK)
using a Pilatus 2M detector
at a sample-detector distance of 4 m and
at a wavelength of λ = 0.1 nm
(I(s) vs s, where s = 4πsinθ/λ, and 2θ is the scattering angle).
One solute concentration of 4.00 mg/ml was measured.
620 successive
3 second frames were collected.
The data were normalized to the intensity of the transmitted beam and radially averaged; the scattering of the solvent-blank was subtracted.
Cell temperature = UNKNOWN. Storage temperature = UNKNOWN |
|