| Synchrotron SAXS 
		
        data from solutions of
        DNA-dependent protein kinase/ X-ray repair cross-complementing protein 5 and 6 complex bound to DNA (DNA-PK monomer/dimer)
		
		    in 
			50 mM Tris-HCl, 100 mM NaCl, 5% glycerol, 0.01% sodium azide, pH 7.5
		
 		were collected
		
		on the
		
			12.3.1 (SIBYLS) beam line  
		
		
			at the Advanced Light Source (ALS) storage ring
		
		
			(Berkeley, CA, USA)
		
		
		
			using a MAR 165 CCD detector
		
		
			at a sample-detector distance of 1.5 m and
		
		
			at a wavelength of λ = 0.103 nm
		
		(I(s) vs s, where s = 4πsinθ/λ, and 2θ is the scattering angle).
                
                In-line size-exclusion chromatography (SEC) SAS was employed. The SEC parameters were as follows: A 50.00 μl sample
                at 5 mg/ml  was injected at a 0.25 ml/min flow rate
                onto a  Shodex KW400 series  column
		
                        at 20°C.
		
			The data were normalized to the intensity of the transmitted beam and radially averaged; the scattering of the solvent-blank was subtracted. 
            
                Storage temperature = UNKNOWN. Number of frames = UNKNOWN
		    
		 |  |