Synchrotron SAXS data from solutions of nsP2 proteins in buffer containing 20 MM Hepes pH7.4, 150 mM NaCl, 1 mM DDT, 5% glycerol was measured at 288 K at the 23A small/wide angle X-ray scattering (SWAXS) endstation of the National Synchrotron Radiation Research Center (Hsinchu, Taiwan). First, 0.1 mL of nsP2 was injected into an HPLC system (Agilent chromatographic system 1260 series) equipped with an online size exclusion column at a flow rate of 0.035 mL/min for SAXS data collection. A 15 keV X-ray beam irradiated the sample flow within a thin-wall (30 μm) quartz capillary cell (2.0 mm diameter) for SAXS measurements (30 s/frame). The sample-to-detector distance (3.42 m) together with a Pilatus 1M-F area detector covers an X-ray momentum transfer q range of 0.01 Å-1 to 0.25 Å-1 (q = 4πλ-1sinθ, where 2θ is the scattering angle, and λ = 0.8266 Å is the wavelength of the X-rays).
|
|