MWexperimental | 70 | kDa |
MWexpected | 70 | kDa |
VPorod | 63 | nm3 |
log I(s)
8.13×10-3
8.13×10-4
8.13×10-5
8.13×10-6
|
s, nm-1
|
|
|
|
|
Synchrotron SAXS
data from solutions of
light-adapted Orange Carotenoid Protein-wtCtag at 0.7 mg/mL
in
50 mM Tris, 150 mM NaCL, pH 7.4
were collected
on the
SWING beam line
at the SOLEIL storage ring
(Saint-Aubin, France)
using a Eiger 4M detector
at a sample-detector distance of 2 m and
at a wavelength of λ = 0.1 nm
(I(s) vs s, where s = 4πsinθ/λ, and 2θ is the scattering angle).
One solute concentration of 0.70 mg/ml was measured
at 21°C.
15 successive
0.990 second frames were collected.
The data were normalized to the intensity of the transmitted beam and radially averaged; the scattering of the solvent-blank was subtracted.
|
|