Dmax unknown – experimental data range validation not possible.
There are no models related to this curve.
Synchrotron SAXS
data from solutions of
Bicelles formed by POPC POPG and CHAPS
in
Tris buffered saline, pH 7.4
were collected
on the
EMBL P12 beam line
at the PETRA III storage ring
(DESY; Hamburg, Germany)
using a Pilatus 6M detector
at a sample-detector distance of 3 m and
at a wavelength of λ = 0.124 nm
(I(s) vs s, where s = 4πsinθ/λ, and 2θ is the scattering angle).
at 20°C.
20 successive
0.100 second frames were collected.
The data were normalized to the intensity of the transmitted beam and radially averaged; the scattering of the solvent-blank was subtracted.
Data was collected on bicelles formed by POPC/POPG (80/20 Mol%) and CHAPS.
The curve shown has a lipid-to-detergent ratio of 5 in Mol%, other ratios are available in the full entry.