Dmax unknown – experimental data range validation not possible.
There are no models related to this curve.
Scattering data from solutions of bovine serum albumin in phosphate buffered saline, pH 7.4 were collected using a AGIPD-1Mpix detector at a sample-detector distance of 3 m and at a wavelength of λ = 0.13 nm (I(s) vs s, where s = 4πsinθ/λ, and 2θ is the scattering angle). The data were normalized to the intensity of the transmitted beam and radially averaged; the scattering of the solvent-blank was subtracted.
BSA solutions were measured on the European XFEL using a "reservoir" to inject the sample into an in-vacuum liquid jet delivery system. The intensity at low angle increases due to imprecise background subtraction (background varying between sample and buffer measurements and/or the presence of aggregates.)