Han Q,
Veríssimo N,
Bryant S,
Martin A,
Huang Y,
Pereira J,
Santos-Ebinuma V,
Zhai J,
Bryant G,
Drummond C,
Greaves T,
Advances in Colloid and Interface Science
:103242
(2024)
DOI
SASDSM9 – Hen egg white lysozyme in 1 mol% ethylammonium nitrate
Synchrotron SAXS data from solutions of hen egg white lysozyme in 1 mol% ethylammonium nitrate, pH 8 were collected on the SAXS/WAXS beam line at the Australian Synchrotron (Melbourne, Australia) using a Pilatus3 S 2M detector at a sample-detector distance of 2.2 m and at a wavelength of λ = 0.1033 nm (I(s) vs s, where s = 4πsinθ/λ, and 2θ is the scattering angle). One solute concentration of 5.00 mg/ml was measured at 25°C. 20 successive 1 second frames were collected. The data were normalized to the intensity of the transmitted beam and radially averaged; the scattering of the solvent-blank was subtracted.