SAXS data from solutions of N-terminal truncated mutant glucuronoyl esterase (S286A, deglycolsylated) in 20 mM sodium acetate, pH 5 were collected using a Xenocs BioXolver L with GeniX3D (University of Copenhagen, Department of Drug Design and Pharmacology) equipped with a Pilatus3 R 300K detector at a sample-detector distance of 0.6 m and at a wavelength of λ = 0.154 nm (I(s) vs s, where s = 4πsinθ/λ, and 2θ is the scattering angle). One solute concentration of 3.60 mg/ml was measured at 21°C. 10 successive 60 second frames were collected. The data were normalized to the intensity of the transmitted beam and radially averaged; the scattering of the solvent-blank was subtracted.
|
|