Synchrotron SAXS data from solutions of X-chromosome linked inhibitor of apoptosis protein (XIAP) in Xiap buffer, pH 7.4 were collected using size-exclusion chromatography SAXS (SEC-SAXS) on the EMBL P12 beam line at PETRA III (Hamburg, Germany) using a Pilatus 2M detector at a sample-detector distance of 3 m and at a wavelength of λ = 0.124 nm (l(s) vs s, where s = 4πsinθ/λ, and 2θ is the scattering angle). SEC-SAXS was performed at 20 °C using the following parameters: Column: Superdex 200 10/300 GL (GE healthcare Life Sciences); Flow rate: 0.5 mL/min; Sample injection concentration: 7.5 mg/mL; Injection volume: 100 μL. The data were collected through the SEC peak of the protein as a series of 15 x 1 second exposures. Each unsubtracted data frame was normalised to the intensity of the transmitted beam and radially averaged and the scattering of an appropriate solvent-blank was subtracted. The resulting subtracted frames were scaled and averaged to generate the final SAXS profile displayed in this entry.
Storage temperature = UNKNOWN. SEC column = UNKNOWN. Sample injection volume = UNKNOWN. Flow rate = UNKNOWN
|
|